EasyManua.ls Logo

GE Kelman DGA900 - Compliance and Type Tests

GE Kelman DGA900
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MA-025 DGA 900 Operator Guide Rev 2.0 12-Apr-19 Page 14 of 92
4 COMPLIANCE
The product is designed to meet the following type tests as listed in Table 4-1.
Table 4-1: Type tests
CATEGORY
STANDARD
CLASS/LEVEL
TEST
EMC Emissions
EN 61326-
1:2006
CISPR 11
A
Radiated & Conducted Emissions
FCC Part 15
Meets the
requirements of A
Radiated & Conducted Emissions
EN 61000-3-2
A
Harmonic Current Emissions
Limits
EMC Immunity
EN 61326-
1:2006
IEC 61000-6-5:
2015
EN 61000-4-2
IV
Electrostatic Discharge
EN 61000-4-3
III
Electromagnetic Field Immunity
EN 61000-4-4
III
Electrical Fast Transients
EN 61000-4-5
III
Surge Immunity
EN 61000-4-6
III
Conducted RF Immunity
EN 61000-4-8
IV & V
Magnetic Field Immunity
EN 61000-4-11
III
Voltage Dips & Interruptions
IEC 61000-4-12
2.5 kV & 1 kV
Oscillatory Wave
IEC 61000-4-16
A
Mains frequency voltage
EN 60255-5
5 kV, 2 kV &
500 V DC
Impulse, Dielectric & Insulation
resistance testing
Environmental
IEC 60068-2-1
−40 °C
Cold
IEC 60068-2-2
55 °C
Dry Heat
IEC 60068-2-6
10 500 Hz,
0.5 g operation
10 500 Hz,
1 g endurance
Vibration
IEC 60068-2-30
55 °C, 95% RH
Damp Heat
EN 60529
IP56
Degree of Protection
Safety
IEC 61010-1
2010
EN 61010-1
2010

Table of Contents