7 MANUAL OIL SAMPLING
A manual oil sample can be taken via the following arrangement.
7.1 Sampling Arrangement
The product has one manual oil sampling port fitted to the base. This consists of a quick-
connect body and a captive body-protector fitting as shown in Figure 7-1. The product
ensures that there is fresh oil at the sampling point when the sampling process takes
place.
Figure 7-1: Oil filter and oil ports (front view of Analysis module)
A male quick-connect fitting and valve assembly is provided in the product installation kit.
GE also recommends the use of a 50-mL ground glass syringe.
Note: The manual oil sampling arrangement and the quick-connect sampling port must
not be cleaned with any type of solvent as this could affect results for subsequent