MA-025 DGA 900 Operator Guide Rev 2.0 12-Apr-19 Page 8 of 92
Advanced high precision Photo-Acoustic Spectroscopy (PAS) sensing technology based
on a robust multi-point individual gas calibration process using certified gases to
calibrate the PGA (Photoacoustic Gas Analyser).
Complete multi-gas analysis (nine gases and moisture) with a user-configurable
sampling frequency from once per hour to once every four weeks including remote alert
capabilities. All gas sensing is carried out internally without a need for carrier gases or
recalibration. Minimal maintenance
*1
and no requirement for costly consumables.
A new ‘Rapid Mode’ allows sampling to be completed in around 30 minutes using five
key gases and is useful in critical situations to give immediate insight on fast-developing
faults.
Improved lower detection limits and measurement repeatability for earlier detection of
imminent faults on the transformer being monitored.
Transformer load tracking (CT analogue input as standard).
Watchdog relay as standard to monitor power usage.
Six configurable relay contacts based on absolute gas and moisture values (with one
dedicated service).
Five optional 4-20 mA configurable analogue input cards.
Three digital inputs (optional configurations available).
Compatible with AC or DC power.
Four sunlight-visible LED arrays on the exterior of the Hub module red for alarm,
yellow for caution (both user-configurable) and green for power, blue for service.
A fully embedded microprocessor with 4 GB non-volatile internal memory storage,
highly scalable analogue and digital I/Os, an embedded webserver, 8 GB SD card to hold
measurement data and an integrated 7 in. colour LCD screen with resistive touch for
simplified local user interaction and visualization of data.
The infrastructure provides a platform for expansion and future feature enhancements
complemented by an extensive range of secure communications options. These include
Ethernet, RS-485, cellular modem (SMS Text Alerts /GPRS), and fibre optic options
(IEC61850 or DNP3). Internal USB connection is provided for commissioning and service,
or local data download. Other options may be available on request.
Compatibility with GE’s Perception transformer fleet management software for data
download, trending and analysis as well as other SCADA systems. The product is a DS-
Agile and Predix-Grid APM ready device with support for industry standard protocols.
*
1
Note: The only recommended maintenance is periodic cleaning of the air filters, in-line
oil filter and battery replacement.