EasyManuals Logo

GE MFE DMS 2 User Manual

GE MFE DMS 2
227 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
5. Working with Stored Data
Page 148 DMS 2 Operating Manual
Step 2:
Activate the STATS Submenu by selecting it with
or so that it’s highlighted.
Step 3:
Access the desired data file, if not already active, by
pressing
below the FILE selection.
Step 4:
Press below the STAT selection until the attribute
you wish to search for is selected. READINGS, EMPTY
POINTS, OBSTRUCTED POINTS, MIN READING, MAX
READING, A-scans, B-scans, and Micro-Grids can be tagged.
Step 5:
Press below the TAG selection until a appears
as shown in Figure 5-7.
NOTE: Multiple attributes can be tagged as part of a given
search. Each attribute must be first identified with the STAT
selection, then tagged with the TAG selection (as described
above). However, not all attributes can be tagged. If an attempt
is made to tag a non-taggable attribute, no check will appear
next to it on the display screen. Refer to Figure 5-7.
Step 6:
After all desired attributes have been tagged, press
below the SEARCH selection to launch a search. In the
SEARCH Mode, the user can navigate through a select set of
data points. Only those points that contain one or more of the
tagged attributes will be displayed while in SEARCH Mode.
Step 7:
To exit the mode, press below the selection titled
QUIT SEARCH, which will only appear once the SEARCH
Mode is activated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MFE DMS 2 and is the answer not in the manual?

GE MFE DMS 2 Specifications

General IconGeneral
BrandGE
ModelMFE DMS 2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals