moisture.IQ User’s Manual 87
Chapter 7. Maintenance
Chapter 7. Maintenance
This chapter covers the following topics:
“Delta F Oxygen Cell Electrolyte” on page 87
“Replacing and Recalibrating Moisture Probes” on page 89
Calibrating the Delta F Oxygen Cell” on page 89
“Delta F Oxygen Cell Electrolyte” on page 87
7.1 Delta F Oxygen Cell Electrolyte
As a result of operating the moisture.IQ, particularly when monitoring dry gases, there may be a gradual loss of
water from the electrolyte. The electrolyte level should be checked at regular intervals to ensure that the cell is
always operating properly. This section describes how to check and replenish the electrolyte in the oxygen cell.
Note: Some applications require that the electrolyte be changed periodically. Consult BHGE.
WARNING! Do not attempt maintenance of the moisture.IQ beyond the instructions in this section. If
you do, you may damage the unit and void the warranty.