1 Introduction
8
Monitor UV-900 User Manual 18-1120-05 Edition AF
1.2 Safety
• The instrument is designed for indoor use only.
• Do not use in a dusty atmosphere or close to spraying water.
IMPORTANT!
Monitor UV-900 is intended for laboratory use only, not for
clinical or in vitro use, or for diagnostic purposes.
WARNING!
When using hazardous chemicals, all suitable protective
measures, such as protective glasses, must be taken.
WARNING
! The instruments must not be opened by the user. They contain
high voltage circuits which can give a lethal electric shock.
WARNING
! The system must be connected to a grounded mains socket.
WARNING
! Always disconnect the power supply before attempting to
replace any item on the instrument.
WARNING!
The instrument uses high intensity ultra-violet light. Do not
disconnect the optical fibers while the monitor is running.
CAUTION!
The flow cell must not be used at pr essur e s above 2 MP a (20 bar,
290 psi). At higher pressures the flow cell may break.
WARNING!
When using hazardous chemicals, take all suitable protective
measures, such as wearing protective glasses and gloves resistant to the
chemicals used. Follow local r egulat ions and in structions for sa fe operati on
and maintenance of the system.
WARNING!
There must always be a sample loop connected to ports 2 and 6
of the injection v alve. This is to prev ent liquid spraying out of the ports when
switching the valv e. This is especially dangerous if hazardous chemicals are
used.
WARNING!
This is a Class A product. In domestic environment this product
may cause radio interference in which the user may be required to take
adequate measur es.