GE MEDICAL SYSTEMS PRECISION RXI SYSTEM PRE-INSTALLATION
D
IRECTION 5123449-100, REVISION 12
_____________________________________________________________________________________________
June 2009 Page 19 / 111
4. Suite/Room Requirements
4.1. Environmental
4.1.1. Relative Humidity and Temperature
RELATIVE HUMIDITY (NON
CONDENSING
TEMPERATURE
STORAGE IN-USE STORAGE IN-USE
COMPONENT
MIN. MAX MIN. MAX. MIN. MAX. MIN. MAX.
Table 30% 70% 30% 70%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
OTS 30% 75% 30% 75%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
Wallstand 30% 75% 30% 75%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
Monitor Suspension
(single and double)
30% 95% 30% 75%
-20°C
-4°F
70°C
140°F
10°C
50°F
40°C
104°F
LCD Monitor 10% 95% 30% 75%
-10°C
14°F
60°C
140°F
10°C
50°F
40°C
104°F
Generator 10% 95% 30% 75%
-20°C
-4°F
70°C
158°F
10°C
50°F
40°C
104°F
Digital 20% 80% 20% 80%
-40°C
-40°F
60°C
18°C
64°F
30°C
86°F
Analog TV chain 30% 75% 30% 75%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
4.1.2. Atmospheric Pressure
ATMOSPHERIC PRESSURE
STORAGE IN-USE
COMPONENT
MIN. MAX. MIN. MAX.
Table - - 700hPa 1060hPa
OTS - - 700hPa 1110hPa
Wallstand - - 700hPa 1060hPa
Monitor Suspension
(single and double)
- - 700hPa 1060hPa
LCD Monitor 500hPa 1060hPa 500hPa 1060hPa
Generator 500hPa 1060hPa 500hPa 1060hPa
Digital 700hPa 1060hPa 700hPa 1060hPa
Analog TV chain 700hPa 1060hPa 700hPa 1060hPa

Table of Contents