GE MEDICAL SYSTEMS PRECISION RXI SYSTEM PRE-INSTALLATION
D
IRECTION 5123449-100, REVISION 12
_____________________________________________________________________________________________
June 2009 Page 19 / 111
4. Suite/Room Requirements
4.1. Environmental
4.1.1. Relative Humidity and Temperature
RELATIVE HUMIDITY (NON
CONDENSING
TEMPERATURE
STORAGE IN-USE STORAGE IN-USE
COMPONENT
MIN. MAX MIN. MAX. MIN. MAX. MIN. MAX.
Table 30% 70% 30% 70%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
OTS 30% 75% 30% 75%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
Wallstand 30% 75% 30% 75%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
Monitor Suspension
(single and double)
30% 95% 30% 75%
-20°C
-4°F
70°C
140°F
10°C
50°F
40°C
104°F
LCD Monitor 10% 95% 30% 75%
-10°C
14°F
60°C
140°F
10°C
50°F
40°C
104°F
Generator 10% 95% 30% 75%
-20°C
-4°F
70°C
158°F
10°C
50°F
40°C
104°F
Digital 20% 80% 20% 80%
-40°C
-40°F
60°C
18°C
64°F
30°C
86°F
Analog TV chain 30% 75% 30% 75%
-20°C
-4°F
60°C
140°F
10°C
50°F
40°C
104°F
4.1.2. Atmospheric Pressure
ATMOSPHERIC PRESSURE
STORAGE IN-USE
COMPONENT
MIN. MAX. MIN. MAX.
Table - - 700hPa 1060hPa
OTS - - 700hPa 1110hPa
Wallstand - - 700hPa 1060hPa
Monitor Suspension
(single and double)
- - 700hPa 1060hPa
LCD Monitor 500hPa 1060hPa 500hPa 1060hPa
Generator 500hPa 1060hPa 500hPa 1060hPa
Digital 700hPa 1060hPa 700hPa 1060hPa
Analog TV chain 700hPa 1060hPa 700hPa 1060hPa

Table of Contents

Related product manuals