GE MEDICAL SYSTEMS PRECISION RXI SYSTEM PRE-INSTALLATION
D
IRECTION 5123449-100, REVISION 12
_____________________________________________________________________________________________
June 2009 Page 87 / 111
11. Seismic Calculations
11.1. Overview
Seismic requirements are determined and specified by the hospital architect of record and may require approval by
the specific state or country agency.
Seismic attachment hardware shown on seismic calculations may differ from hardware supplied with system. Any
additional hardware that is required will be the responsibility of the institution and/or their contractor. Contact your
Project Manager Installations with any related questions.
Seismic calculations included in this chapter are per California Building Code.
11.2. Calculations
Calculations are included for the following:
• 9.2.1 Elevating Table
o 9.2.1.1 Slab on Grade
o 9.2.1.2 Upper Floor
• 9.2.2 Non-Elevating Table
o 9.2.2.1 Slab on Grade
o 9.2.2.2 Upper Floor
• 9.2.3 Generator Cabinet
o 9.2.3.1 Slab on Grade
o 9.2.3.2 Upper Floor
• 9.2.4 Integrated Console
o 9.2.4.1 Slab on Grade
o 9.2.4.2 Upper Floor
• 9.2.5 Table Cabinet
o 9.2.5.1 Slab on Grade
o 9.2.5.2 Upper Floor
• 9.2.6 Monitor Suspension
• 9.2.7 Non-tilting Wallstand
o 9.2.7.1 Slab on Grade
o 9.2.7.2 Upper Floor
• 9.2.8 Tilting Wallstand
o 9.2.8.1 Slab on Grade
o 9.2.8.2 Upper Floor
• 9.2.9 Overhead Tube Suspension