GE MEDICAL SYSTEMS PRECISION RXI SYSTEM PRE-INSTALLATION
D
IRECTION 5123449-100, REVISION 12
_____________________________________________________________________________________________
Page 6 / 111 June 2009
REVISION HISTORY
REV DATE REASON FOR CHANGE
1 November 2004 Initial Release
2 June 2005
New OTS specs, table base plate change, cable lengths change, additional
wallstand and pos cabinet mounting info, added floor flatness requirement,
seismic information added, China cable requirements added
3 February 2006 Table base plate information corrections.
4 March 2006
Corrections and additional tube angulations information.
Precision RXi Air Kerma Scatter Rates.
Broadband information for Europe
5 July 2006 Positioner tabletop longitudinal displacement; room layout.
6 September 2006 Generator 50 kW deleted, moving the Bucky, room layout.
7 April 2007 Console joystick protection and new touch screen 15”
8 November 207 New monitor, new dual monitor cart.
9 April 2008 Replacement monitor cart
10 September 2008
Maximum momentary power demand kVA becomes
Minimum recommended distribution transformer rating
Page 68
11 January 2009 Review of electrical safety direction
12 June 2009 New monitor suspension