GE MEDICAL SYSTEMS PRECISION RXI SYSTEM PRE-INSTALLATION
D
IRECTION 5123449-100, REVISION 12
_____________________________________________________________________________________________
June 2009 Page 65 / 111
With GE Industrial Panel
CB
CB
CB
3 winding
Isolation XFR
(UL506 Std)
EO
EO
Control
Generator
3φ-480V
60Hz
480V
7.5KVA
240V/120V
Hospital Power
3φ/480V
• Has Emergency Off (EO)
CB
Positioner Cabinet
1φ-230
Remote
Positioner
Controller
Table
OTS
1φ-230V
CB
230V
230V
230V
CB
Digital Imager
(PlatinumOne)
Isolation Transformer
1φ-230V
50/60Hz
Control Room
monitor
GIM
1-2 exam room monitors
(Digital system only)
Wallstand/Bucky
TV chain and II
1-2 exam room monitors
(Analog system only)
UL508A
Industrial
Standard
Panel
GE
supplied
Power requirements with industrial panel (North
America – 480V, 60HZ)
CB
CB
CB
3 winding
Isolation XFR
(UL506 Std)
EO
EO
Control
Generator
3φ-480V
60Hz
480V
7.5KVA
240V/120V
Hospital Power
3φ/480V
• Has Emergency Off (EO)
CB
Positioner Cabinet
1φ-230
Remote
Positioner
Controller
Table
OTS
1φ-230V
CB
230V
230V
230V
CB
Digital Imager
(PlatinumOne)
Isolation Transformer
1φ-230V
50/60Hz
Control Room
monitor
GIM
1-2 exam room monitors
(Digital system only)
Wallstand/Bucky
TV chain and II
1-2 exam room monitors
(Analog system only)
UL508A
Industrial
Standard
Panel
GE
supplied
CBCB
CBCB
CBCB
3 winding
Isolation XFR
(UL506 Std)
3 winding
Isolation XFR
(UL506 Std)
EOEO
EOEO
ControlControl
Generator
3φ-480V
60Hz
480V
7.5KVA
240V/120V
Hospital Power
3φ/480V
• Has Emergency Off (EO)
CBCB
Positioner Cabinet
1φ-230
Remote
Positioner
Controller
Table
OTS
1φ-230V
CBCB
230V
230V
230V
CBCB
Digital Imager
(PlatinumOne)
Isolation Transformer
1φ-230V
50/60Hz
Control Room
monitor
GIM
1-2 exam room monitors
(Digital system only)
Wallstand/Bucky
TV chain and II
1-2 exam room monitors
(Analog system only)
UL508A
Industrial
Standard
Panel
GE
supplied
Power requirements with industrial panel (North
America – 480V, 60HZ)
CB
CB
3 winding
Isolation XFR
(UL506 Std)
EO
EO
Control
400V
7.5KVA
240V/120V
Hospital Power
3φ/400V
• Has Emergency Off (EO)
230V
230V
230V
CB
Generator
3φ-400V
50Hz
CB
Positioner Cabinet
1φ-230
Remote
Positioner
Controller
Table
OTS
1φ-230V
CB
CB
Digital Imager
(PlatinumOne)
Isolation Transformer
1φ-230V
50/60Hz
Control Room
monitor
GIM
1-2 exam room monitors
(Digital system only)
Wallstand/Bucky
TV chain and II
1-2 exam room monitors
(Analog system only)
CE
Industrial
Panel
Standard
GE
supplied
Power requirements with industrial panel
(Europe – 380-400V, 50HZ)
CB
CB
3 winding
Isolation XFR
(UL506 Std)
EO
EO
Control
400V
7.5KVA
240V/120V
Hospital Power
3φ/400V
• Has Emergency Off (EO)
230V
230V
230V
CB
Generator
3φ-400V
50Hz
CB
Positioner Cabinet
1φ-230
Remote
Positioner
Controller
Table
OTS
1φ-230V
CB
CB
Digital Imager
(PlatinumOne)
Isolation Transformer
1φ-230V
50/60Hz
Control Room
monitor
GIM
1-2 exam room monitors
(Digital system only)
Wallstand/Bucky
TV chain and II
1-2 exam room monitors
(Analog system only)
CE
Industrial
Panel
Standard
GE
supplied
CBCB
CBCB
3 winding
Isolation XFR
(UL506 Std)
3 winding
Isolation XFR
(UL506 Std)
EOEO
EOEO
ControlControl
400V
7.5KVA
240V/120V
Hospital Power
3φ/400V
• Has Emergency Off (EO)
230V
230V
230V
CBCB
Generator
3φ-400V
50Hz
CBCB
Positioner Cabinet
1φ-230
Remote
Positioner
Controller
Table
OTS
1φ-230V
CBCB
CBCB
Digital Imager
(PlatinumOne)
Isolation Transformer
1φ-230V
50/60Hz
Control Room
monitor
GIM
1-2 exam room monitors
(Digital system only)
Wallstand/Bucky
TV chain and II
1-2 exam room monitors
(Analog system only)
CE
Industrial
Panel
Standard
GE
supplied
Power requirements with industrial panel
(Europe – 380-400V, 50HZ)