GE MEDICAL SYSTEMS PRECISION RXI SYSTEM PRE-INSTALLATION
D
IRECTION 5123449-100, REVISION 12
_____________________________________________________________________________________________
Page 72 / 111 June 2009
CURRENT
REQUIREMENT – US System
Precision RXi
COMPONENT
CURRENT
MOMENTARY CONTINUOUS
Generator – 65kW 105A 5A(standby current)
Generator – 80kW 130A 5A(standby current)
Positioner N/A 17.4A
Digital Transformer N/A 4.4 A
OTS N/A 1.3A
CURRENT
REQUIREMENT – European System
Precision RXi
COMPONENT
CURRENT
MOMENTARY CONTINUOUS
Generator – 65kW 125A 5A(standby current)
Generator – 80kW 155A 5A(standby current)
Positioner N/A 17.4 A
Digital Transformer N/A 4.4 A
65 kW GENERATOR POWER DEMAND – US Systems
Continuous power demand 2.4 kVA at 480V, 50/60 Hz
Minimum recommended distribution transformer rating 85 kVA at 480V, 50/60 Hz
Maximum momentary mA 800mA at 480V, 50/60 Hz
Maximum momentary kVp 120kVp at 480V, 50/60 Hz
65 kW GENERATOR POWER DEMAND – European Systems
Continuous power demand 2.0 kVA at 400V, 50/60 Hz
Minimum recommended distribution transformer rating 85 kVA at 400V, 50/60 Hz
Maximum momentary mA 800mA at 400V, 50/60 Hz
Maximum momentary kVp 120kVp at 400V, 50/60 Hz
80 kW GENERATOR POWER DEMAND – US Systems
Continuous power demand 2.4 kVA at 480V, 50/60 Hz
Minimum recommended distribution transformer rating 105 kVA at 480V, 50/60 Hz
Maximum momentary mA 1000mA at 480V, 50/60 Hz
Maximum momentary kVp 120kVp at 480V, 50/60 Hz
80 kW GENERATOR POWER DEMAND – European Systems
Continuous power demand 2.0 kVA at 400V, 50/60 Hz
Minimum recommended distribution transformer rating 105 kVA at 400V, 50/60 Hz
Maximum momentary mA 1000mA at 400V, 50/60 Hz
Maximum momentary kVp 120kVp at 400V, 50/60 Hz