PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
ix
ELECTROMAGNETIC COMPATIBILITY (EMC) (CONT.)
Recommended Separation Distances for Portable and Mobile RF Communications
Equipment and the Proteus XR/a system
Frequency of
Transmitter
150KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Equation
d= 1.2
d= 1.2
d= 2.3
Rated Power of
Transmitter
(W)
DISTANCE
(meters)
DISTANCE
(meters)
DISTANCE
(meters)
0.01
0.12 0.12 0.23
0.1
0.38 0.38 0.73
1
1.2 1.2 2.3
10
3.8 3.8 7.3
100
12 12 23
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the
equation in the corresponding column, where P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note: These are guidelines. Actual conditions may vary.