CHAPTER 13 PHYSICAL REQUIREMENTS OF ROOM
13-1 Environmental Requirements/Limitations
13-1-1 Room Climate
Relative humidity and temperature
See Table 13-1. To obtain relative humidity and temperature requirements for components
TABLE 13-1
CLIMATE REQUIREMENT (BY COMPONENT) - RELATIVE HUMIDITY AND TEMPERATURE
PRODUCT OR RELATIVE HUMIDITY TEMPERATURE
COMPONENT
IN-USE STORAGE IN-USE STORAGE
MIN. MAX. MIN. MAX. MIN. MAX. MIN. MAX.
Elevating table 10% 75% 5% 95%
+10°C
40°C
-20°C
+70°C
OTS 10% 75% 5% 95%
+10°C
40°C
-20°C
+70°C
Generator 10% 75% 5% 95%
+10°C
40°C
-20°C
+70°C
System Console 10% 75% 5% 95%
+10°C
40°C
-20°C
+70°C
Tomolink Option 10% 75% 5% 95%
+10°C
40°C
-20°C
+70°C
X-ray Tube (Rad 14) 10% 75% 5% 95%
+10°C
40°C
-20°C
+70°C
X-ray Tube (MX 100) 10% 80% 5% 95%
0°C
40°C
-40°C
+80°C
SG120 Wallstand 20% 85% 10% 95%
+10°C
40°C
-20°C
+70°C