PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
12-2
12-2-2 Operator Errors
Operator errors are detailed in Table 12-1.
Errors may be due to:
Inadequate parameter selection
Inadequate AEC cell choice
Collimator blades excessively closed or excessively open
Premature release of exposure switch
Door of x-ray room open
The consequences of an operator error detected by the Proteus XR/a Generator
are:
Equipment inhibited, making exposures impossible
Warning light and buzzer activated
What to do step by step:
Find the root cause by referring to Table 12-1.
Press the reset Key
Make one attempt to continue the exam
The faulty operation should disappear.