PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
14-2
14- 2 Table Specifications
See Table 14-1 for Table Specifications.
See Table 14-2 for Table Bucky Specifications.
TABLE 14-1
TABLE SPECIFICATIONS
Parameter Elevating Tables
Table Top Length 2174.5 mm
Table Top Width 685.5 mm
Height
500±10mm ~ 800±10mm Elevating
Table Top Longitudinal Travel
+/-390±10mm
Table Top Transverse Travel
+/-116±10mm
Table Top Operating Force <49N
Table Top Detent Force >150N
Vertical Travel Speed (50Hz)
42±5mm/s
Vertical Travel Speed (60Hz)
52±5mm/s
Travel Stopping Distance <10mm
Down motion prohibit Force
20±2kg
Table top Load (Max.) 220 kg
Table Top Filtration <1.0 mm Aluminium equivalency at 100 kVp
Table Top to Film Distance (For Bucky) <=76mm
Table Weight 250 kg
Power Supply 220+/-10%VAC, 50Hz/60Hz, 5Amp
TABLE 14-2
TABLE BUCKY SPECIFICATIONS
Parameter Specification
Longitudinal Travel in Table
590±10mm
Grid Size
438mm×479.5mm
Grid
36line/cm; 12:1 ratio; 100 cm focus
Cassette Size Sensing
18 cm (7 in.) minimum to 43 cm (17 in.) maximum