Components and Functions (Theory)
5-18 Vivid E80/E90/E95Service Manual
GC091052
Rev. 3
Description of the Locking Mechanism (continued)
The micro switch activation is used by the Motor Controller to
detect when the Operating Panel is in locked position.
The XY (frogleg) brakes will stay ON if the Vivid E80/E90/E95 is
powered down when the Operating Panel is locked. This
enables release of the Park Lock, initiated by the micro switches
on the Front Handle of the Operating Panel.