System Setup
3-2 Vivid E80/E90/E95Service Manual
GC091052
Rev. 3
Setup reminders
Setup warnings
DANGER
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE
OF OPENING THE AC GROUND LINE (I.E., METER’S
GROUND SWITCH IS OPEN), DON’T TOUCH THE
ULTRASOUND SYSTEM!
CAUTION
To prevent electrical shock, connect the Ultrasound system to a
properly grounded power outlet. Do not use a three to two
prong adapter. This defeats safety grounding.
CAUTION
Do not wear the ESD wrist strap when you work on live circuits
and more than 30 V peak is present.
CAUTION
Do not operate this Ultrasound system unless all board covers
and frame panels are securely in place. System performance
and cooling require this.