Replacement Procedures
8-78 Vivid E80/E90/E95 – Service Manual
GC091052
Rev. 3
Install the Front Casters
Follow this procedure to install the front casters:
1. Position the front caster so it align with the fastening screws.
2. Install the three fixing screws (M8x20 mm), Torque: 20.5 Nm
(15.1 lb.-ft.).
3. Remove the wooden wedge.
4. Roll the Ultrasound system off the bevel edged board.
5. To replace the other Front Caster, repeat all the steps from
‘Remove the Front Casters’ on page 8-77, but now using the
bevel edged board and the wooden wedge on the other side
of the Ultrasound system.
6. Install the foot rest bumper.
7. Install the two side covers.