Site Preparations
2-6 Vivid E80/E90/E95Service Manual
GC091052
Rev. 3
Power plug
If the Ultrasound system arrives without a power plug, or with
the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
Power stability requirements
Voltage drop-out:
Max 10 ms.
Power transients (all applications):
Less than 25% of nominal peak voltage for less than 1
millisecond for any type of transient, including line frequency,
synchronous, asynchronous, or aperiodic transients.