HT Series II Inert Gas Purge
10-1474 Issue 3-6 March 2014 5
Combustible solvents
Warning: Risk of vapour ignition. Only operate the inert gas purge system in a
ventilated environment and consider safety when evaporating any combustible
solvents. Genevac's position regarding the evaporation of such solvents,
particularly with respect to the European ATEX directive, is available on our
website, at www.genevac.com or from your local Genevac distributor.
Electrical earthing
Warning: Risk of electric shock. This equipment must be earthed. This is a
safety class I product according to IEC classification. It must never be used with
any interruption to the safety earth conductor. It is an installation class II product
and is intended to operate from a normal single-phase supply.
Any maintenance or repair of this product, not detailed in this manual, must be carried out by
Genevac personnel (or approved representatives of Genevac) using only Genevac approved
spare parts.
Limitation of Use
The inert gas purge system is designed to be used in conjunction with the a Genevac
HT Series II evaporation system with which it was supplied. It must not be transferred to
another system.