Zoom80 | 88Check & Adjust
5. MEAS (F1) to measure and to continue to the next screen.
Instruments change automatically
to the other face.
The fine pointing has to be
performed manually in both faces.
6. TOOLS Combined II
MEAS (F1) to measure the same target in the other face and to
calculate the instrument errors.
If one or more errors are bigger than the predefined limits, the proce-
dure has to be repeated. All measurements of the current run are
rejected and none of them is averaged with the results from previous
runs.
Step Description
180°
180°