Zoom80 | 94Check & Adjust
6. TOOLS T-Axis Adjustment Accuracy
<No.of Meas:> Shows the number of runs executed. One run
consists of a measurement in face I and face II.
< a T-axis:> shows the standard deviation of the determined tilting
axis error. The standard deviation can be calculated from the second
run onwards.
It is recommended to measure at least two runs.
7. MEAS (F5) if more runs have to be added. Continue with step 3.
OR
CONT (F1) to accept the measurements and to proceed to TOOLS
T-Axis Adjustment Result. No more runs can be added later.