Troubleshooting
6
Maquet Orchide
IFU 04661 EN 02
33 / 44
6 Troubleshooting
Problem Likely cause Corrective action
The sterilisable handle does
not click into place correctly
The locking mechanism is
damaged
Replace the handle
No image after starting the
camera
The camera is defective Replace the camera
The monitor is defective Replace the monitor
Other reason Contact the Getinge technical
department
No image after setting up the
OHDII VP01 QL FHD or OHDII
VP01 QL+ FHD camera
The camera is not on a light-
head pre-wired for video for a
wired camera.
Install the camera on a light-
head pre-wired for video.
No image after starting the
OHDII AIR03 QL FHD camera
Communication problem
1. Check on the touchscreen
control panel whether the
camera is on
2. Disconnect and reconnect
the receiver power cable
3. Re-register the camera
4. Contact the Getinge tech-
nical department
Image lost for more than
20seconds with the OHDII
AIR03 QL FHD camera
Interference with another sys-
tem
1. The system should reset it-
self within around
20seconds
2. Modify the zoom level
3. Contact the Getinge tech-
nical department
Tab.9: Troubleshooting and mechanical failures