Technical specifications
EMC declaration
9
Maquet Orchide
IFU 04661 EN 02
41 / 44
9.2 EMC declaration
CAUTION!
Risk of malfunction of the device
If the device is used in conjunction with other equipment, its operation and
performance may be affected.
Do not use the device alongside other equipment or stacked with other
equipment except after observing the normal operation of the device and the
other equipment.
CAUTION!
Risk of malfunction of the device
The use of hand-held RF communications equipment (including antenna
cables and external antennas) alongside the device or specified cables may
affect the operation and performance of the device.
Do not use hand-held RF communications equipment at within 30cm of the
device.
CAUTION!
Risk of malfunction of the device
The use of accessories, transducers or cables other than those supplied or
recommended by the manufacturer of this device may cause increased elec-
tromagnetic emissions or a decreased immunity of this device, and may res-
ult in improper operation.
Use only accessories and cables supplied or specified by the manufacturer.
NOTE
Electromagnetic interference may result in temporary extinction or temporary flick-
ering of the light, which will resume its initial operation once the interference has
ceased.
Type of test Test methods Range of fre-
quencies
Boundaries
Measurement of conducted
emissions on the main ports
EN 55011 GR1 CL
A
3
0.15 - 0.5MHz 66 dBµV - 56 dBµV
QP
56 dBµV - 46 dBµV A
0.5 - 5MHz 56 dBµV PQ
46 dBµV A
5 - 30MHz 60 dBµV PQ
50 dBµV A
Tab.17: EMC declaration
3
The emission characteristics of this device enable it to be used in industrial areas and hos-
pital settings (Class A as defined in CISPR 11). If used in a residential environment (for which
class B defined in CISPR 11 is normally required), this device may not provide sufficient pro-
tection for radio frequency communication services. The user may need to take corrective
measures, such as relocating or re-orienting the device.