GOM-804 & GOM-805 使用手册
106
BINNing 指令
Binning commands are only applicable to GOM-805.
BINNing:COUNt:CLEar
Set
Query
Description
Clear all bin sorting function test result counts.
Syntax
BINNing:COUNt:CLEar
Parameter/
<None>
BINNing:COUNt:TOTal
Query
Description
Returns the total number (count total) of test bin
results.
Query Syntax
BINNing:COUNt:TOTal?
Return parameter
<NR1>
0~999999999
Example
BINN:COUN:TOT?
>150
Indicates that the total number (count total) of test results
(pass and fail) is 150.
BINNing:COUNt:OUT
Query
Description
Returns the number of failed (judged OUT) test results
for the bin sorting function test.
Query Syntax
BINNing:COUNt:OUT?
Return parameter
<NR1>
0~99999999
Example
BINN:COUN:OUT?
>50
Indicates that the number of failed test results is 50.
BINNing<X>:COUNt:RESult
Query
Description
Returns the number of passed (judged IN) test results
for the selected bin.