GOM-804 & GOM-805 使用手册
107
Query Syntax
BINNing<X>:COUNt:RESult?
Parameter
<X>
1~8
Return parameter
<NR1>
0~99999999
Example
BINN1:COUN:RES?
>100
Indicates that bin1 has a pass count of 100.
BINNing<X>:LIMit:LOWer
Set
Query
Description
Sets or returns the lower limit value (absolute value)
for the selected bin.
Syntax
BINNing<X>:LIMit:LOWer {<NRf>[,<String>]}
BINNing<X>:LIMit:LOWer?
Parameter
<X>
1~8
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0000~999.9999E±X
Example
BINN1:LIM:LOW 23.8,kohm
Sets the bin1 lower limit value to 23.8kΩ.
BINN1:LIM:LOW?
>23.8000E+3
Returns the lower limit as 23.8kΩ.
BINNing<X>:LIMit:UPPer
Set
Query
Description
Sets or returns the upper limit value (absolute value)
for the selected bin.
Syntax
BINNing<X>:LIMit:UPPer {<NRf>[,<String>]}
BINNing<X>:LIMit:UPPer?
Parameter
<X>
1~8
<NRf>
000.0000~999.9999

Table of Contents