GOM-804 & GOM-805 使用手册
109
Example
BINN1:PERC:UPP 150.95
Sets the bin1 upper limit percent value to +150.95%.
BINN1:PERC:UPP?
>150.95
Returns the upper limit percentage value as +150.95%.
BINNing:LIMit:BEEPer
Set
Query
Description
Sets or returns beeper mode for the bin sorting
function.
Syntax
BINNing:LIMit:BEEPer {OFF|PASS|FAIL}
BINNing:LIMit:BEEPer?
Parameter/
Return parameter
OFF
Turns the beeper off.
PASS
The beeper will sound on a pass test
result.
FAIL
The beeper will sound on a fail test
result.
Example
BINN:LIM:BEEP OFF
Turns the beeper off.
BINNing:LIMit:DISP
Set
Query
Description
Sets or returns the bin sorting function display mode.
Syntax
BINNing:LIMit:DISP {COMP|COUNT}
BINNing:LIMit:DISP?
Parameter/
Return parameter
COMP
The display is set to compare mode.
COUNT
The display is set to count mode.
Example
BINN:LIM:DISP COMP
Sets the bin sorting function display mode to compare.
BINNing:LIMit:MODE
Set
Query
Description
Sets or returns the setting mode for upper and lower
limits (absolute or Δ%).
Syntax
BINNing:LIMit:MODE {ABS|DPER}
BINNing:LIMit:MODE?

Table of Contents