GOM-804 & GOM-805 使用手册
108
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0000~999.9999E±X
Example
BINN1:LIM:UPP 0.95,maohm
Sets bin1 upper limit value to 0.95MΩ.
BINN1:LIM:UPP?
>0.9500E+6
Returns the upper limt as 0.95MΩ.
BINNing<X>:PERCent:LOWer
Set
Query
Description
Sets or returns the lower value percentage value for the
selected bin. The value is a percentage offset from the
reference value.
BINNing<X>:PERCent:LOWer <NRf>
BINNing<X>:PERCent:LOWer?
Parameter
<X>
1~8
<NRf>
000.00~999.99
Return parameter
<NR2>
000.00~999.99
Example
BINN1:PERC:LOW 10.15
Sets the bin1 lower limit percent value to -10.15%.
BINN1:PERC:LOW?
>10.15
Returns the lower limit percentage value as -10.15%.
BINNing<X>:PERCent:UPPer
Set
Query
Description
Sets or returns the upper value percentage value for the
selected bin. The value is a percentage offset from the
reference value.
BINNing<X>:PERCent:UPPer <NRf>
BINNing<X>:PERCent:UPPer?
Parameter
<X>
1~8
<NRf>
000.00~999.99
Return parameter
<NR2>
000.00~999.99

Table of Contents