GOM-804 & GOM-805 使用手册
1 41
Parameter
<NRf>
-50.0~399.9 (Unit: ºC)
Return parameter
<NR2>
-50.0~399.9 (Unit: ºC)
Example
TEMP:CONV:TEMP 25.6
Sets the initial temperature to +25.6ºC.
TEMPerature:DATa
Query
Description
Returns the PT-100 sensor temperature measurement in
degrees Celsius.
Query Syntax
TEMPerature:DATa?
Return parameter
<NR3>
-50.0~399.9
Example
TEMP:DAT?
>0.250E+2
Returns the temperature as 25ºC.
TEMPerature:STATe
Set
Query
Description
Sets or returns the temperature function state.
Syntax
TEMPerature:STATe {<NR1>|OFF|ON}
TEMPerature:STATe?
Parameter/
Return parameter
<NR1>
0:OFF
1:ON
OFF
Turn the temp function off.
ON
Turn the temp function on.
Example
TEMP:STAT ON
Sets the temp function on.
TEMPerature:UNIT
Set
Query
Description
Sets or returns the temperature unit. (Only used for the
display readback.)
Syntax
TEMPerature:UNIT {DEGC|DEGF}
TEMPerature:UNIT?
Parameter/
Return parameter
DEGC
ºC
DEGF
ºF

Table of Contents