GOM-804 & GOM-805 使用手册
1 40
Example
TEMP:CONV:DISP 1
Sets the temperature display mode for the temperature
conversion function is ΔT.
TEMPerature:CONVersion:MATH:DATa
Query
Description
Returns conversion function deviation value.
Query Syntax
TEMPerature:CONVersion:MATH:DATa?
Return parameter
<NR3>
±0.000~9.999E±X
Example
TEMP:CONV:MATH:DAT?
Returns 1.250E+2.
TEMPerature:CONVersion:RESistance
Set
Query
Description
Sets or returns the initial resistance for the temperature
conversion function.
TEMPerature:CONVersion:RESistance
{<NRf>[,<String>]}
TEMPerature:CONVersion:RESistance?
Parameter
<NRf>
000.0001~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0001~999.9999E±X
Example
TEMP:CONV:RES 10.00,maohm
Sets initial resistance value to 10.00MΩ.
TEMP:CONV:RES?
>10.0000E+6
Returns the initial resistance as 10.00MΩ.
TEMPerature:CONVersion:TEMPerature
Set
Query
Description
Sets or returns the initial temperature for the
temperature conversion function.
TEMPerature:CONVersion:TEMPerature <NRf>
TEMPerature:CONVersion:TEMPerature?

Table of Contents