GOM-804 & GOM-805 使用手册
1 32
SYSTem:AVERage:DATa
Set
Query
Description
Sets or returns the number of measurements used for
the average function.
SYSTem:AVERage:DATa <NR1>
SYSTem:AVERage:DATa?
Parameter/
Return parameter
<NR1>
2~10
Example
SYST:AVER:DAT 5
5 measurements are used to perform the average function.
SYSTem:AVERage:STATe
Set
Query
Description
Sets or returns the average function state.
SYSTem:AVERage:STATe <NR1> | {OFF|ON}
SYSTem:AVERage:STATe?
<NR1>
0:OFF.
1:ON.
OFF
Turn the average function off.
ON
Turn the average function on.
Example
SYST:AVER:STAT OFF
Turns the average function off.
SYSTem:BRIGhtness
Set
Query
Description
Sets or returns the brightness level.
SYSTem:BRIGhtness <NR1>
SYSTem:BRIGhtness?
Parameter/
Return parameter
<NR1>
1(dim)~5(bright)
Example
SYST:BRIG 4
Turns the brightness level to 4.