EasyManua.ls Logo

GW Instek GOM-804 - Binning:limit:reference; Binning:limit:result

GW Instek GOM-804
164 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GOM-804 & GOM-805 使用手册
110
Parameter/
Return parameter
ABS
The test results are judged from
absolute values.
DPER
The test results are judged from a
reference value ± a percentage offset.
(delta percent)
Example
BINN:LIM:MODE DPER
Sets the mode to Δ%.
BINNing:LIMit:REFerence
Set
Query
Description
Sets or returns the limit reference value for the bin
sorting function.
BINNing:LIMit:REFerence {<NRf>[,<String>]}
BINNing:LIMit:REFerence?
Parameter
<NRf>
000.0001~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0001~999.9999E±X
Example
BINN:LIM:REF 100
Sets the limit reference value to 100Ω.
BINN:LIM:REF?
>100.0000E+0
Returns the reference as 100Ω.
BINNing:LIMit:RESult
Query
Description
Returns the bin sorting function test result.
Query Syntax
BINNing:LIMit:RESult?
Return parameter
<NR1>
1~8: Bin1~Bin8
9: Bin Out
Example
BINN:LIM:RES?
>1
Indicates a pass for bin1.

Table of Contents

Other manuals for GW Instek GOM-804

Related product manuals