GOM-804 & GOM-805 使用手册
118
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0001~999.9999E±X
Example
CALC:COMP:LIM:REF 10.00,mohm
Sets the limit reference value to 10.00mΩ.
CALC:COMP:LIM:REF?
>10.0000E-3
Returns the limit as 10.00mΩ.
CALCulate:COMPare:LIMit:RESult
Query
Description
Returns the compare function test result.
Query Syntax
CALCulate:COMPare:LIMit:RESult?
Return parameter
<NR1>
0: LO
1: IN
2: HI
Example
CALC:COMP:LIM:RES?
>2
Indicates that the test result is HI.
CALCulate:COMPare:LIMit:UPPer
Set
Query
Description
Sets or returns the upper limit value for the compare
function.
CALCulate:COMPare:LIMit:UPPer {<NRf>[,<String>]}
CALCulate:COMPare:LIMit:UPPer?
Parameter
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0000~999.9999E±X