GOM-804 & GOM-805 使用手册
121
Return parameter
<NR3>
000.0000~999.9999E±X
Example
CALC:SCAN:LIM:LOW 0.123,maohm
Sets the lower limit value to 0.123MΩ.
CALC:SCAN:LIM:LOW?
>0.1230E+6
Returns the lower limit as 0.123MΩ.
CALCulate:SCAN:LIMit:MODE
Set
Query
Description
Sets or returns the scan function compare mode.
Syntax
CALCulate:SCAN:LIMit:MODE {ABS|DPER}
CALCulate:SCAN:LIMit:MODE?
Parameter/
Return parameter
ABS
The test results are judged from
absolute values.
DPER
The test results are judged from a
reference value ± a percentage offset.
(delta percent)
Example
CALC:SCAN:LIM:MODE ABS
Sets compare mode to absolute values.
CALCulate:SCAN:LIMit:REFerence
Set
Query
Description
Sets or returns the reference limit for the scan function.
Syntax
CALCulate:SCAN:LIMit:REFerence {<NRf>[,<String>]}
CALCulate:SCAN:LIMit:REFerence?
Parameter
<NRf>
000.0001~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set,the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0001~999.9999E±X
Example
CALC:SCAN:LIM:REF 10.00,mohm
Sets the reference limit to 10.00mΩ.
CALC:SCAN:LIM:REF?
>10.0000E-3
Returns the reference limit as 10.00mΩ.