GOM-804 & GOM-805 使用手册
122
CALCulate:SCAN:LIMit:UPPer
Set
Query
Description
Sets or returns upper limit of the scan function.
CALCulate:SCAN:LIMit:UPPer {<NRf>[,<String>]}
CALCulate:SCAN:LIMit:UPPer?
Parameter
<NRf>
000.0000~999.9999
<String>
mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present
range.
Return parameter
<NR3>
000.0000~999.9999E±X
Example
CALC:SCAN:LIM:UPP 1.37,kohm
Sets the upper limit to 1.37kΩ.
CALC:SCAN:LIM:UPP?
>1.3700E+3
Returns the upper limit as 1.37kΩ.
CALCulate:SCAN:PERCent:LOWer
Set
Query
Description
Sets or returns lower limit percent value for the scan
function.
CALCulate:SCAN:PERCent:LOWer <NRf>
CALCulate:SCAN:PERCent:LOWer?
Parameter
<NRf>
000.00~999.99
Return parameter
<NR2>
000.00~999.99
Example
CALC:SCAN:PERC:LOW 10.00
Sets the lower limit percent value to -10.00%.
CALC:SCAN:PERC:LOW?
>10.00
Returns the lower limit as -10.00%.
CALCulate:SCAN:PERCent:UPPer
Set
Query
Description
Sets or returns the upper limit percent value for the
scan function.