GOM-804 & GOM-805 使用手册
1 46
USERdefine<X>:ACTive
Set
Query
Description
Sets or returns the active output state of the selected
Userdefine pin.
USERdefine<X>:ACTive <NR1>
USERdefine<X>:ACTive?
Parameter/
Return parameter
<X>
Userdefine pin 1~2
<NR1>
1:active low state
2:active high state
Example
USER1:ACT 1
Sets the userdefine1 pin IO to active low state.
USERdefine<X>:FIRStdata
Set
Query
Description
Sets or returns the first operand for the selected user
define pin.
USERdefine<X>:FIRStdata <NR1>
USERdefine<X>:FIRStdata?
Parameter/ Return
parameter
<X>
Userdefine pin 1~2
<NR1>
1~8:bin1~bin8 state
9:bin out state
10:hi state
11:low state
12:pass state
13:fail state
Example
USER1:FIRS 12
Sets first operand of userdefine1 as pass state.
USERdefine<X>:LOGic
Set
Query
Description
Sets or returns operator for the selected user define pin.