EasyManua.ls Logo

GW Instek LCR-6100 - Accuracy; Measurement Accuracy

GW Instek LCR-6100
143 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Examples
88
Step 1. Turn the LCR-6000 Series ON.
Step 2. Set up the measurement conditions by filling in
the fields on the MEAS DISPLAY page.
Move to the FUNC field using the cursor
keys and choose Cs-D.
Move to the FREQ field using the cursor
keys and input 1 kHz.
Move to the LEVEL field using the cursor
keys and input 1 V
Step 3. Connect the test fixture to the LCR-6000
Series.
Step 4. To compensate for the test fixture, a user
correction is required.
Don’t connect any DUT to the unit, as shown
below:
1. Press the [Measure] key and the [OPEN
SHORT] soft-key.
2. Move to the OPEN TEST or SPOT field
using the cursor keys.
3. Don’t connect the clips to anything, as
shown here:
4. Press the [MEAS OPEN] soft-key and
then the [OK] soft-key. Wait until the
“Correction finished” message is
displayed.
5. Press the [ON] soft-key if the OPEN field
is set to OFF.

Table of Contents

Related product manuals