LCR-8000G Series User Manual 
118 
Create a new multi-step program. 
Example: :multi:new demo (file name demo) 
Query the results of the test for each step. 
Parameter: 0 (Pass), 1 (Fail Hi), 2 (Fail Lo) 
Example: 1, +1.5E-7, 0, –0.2E-4 (step 1 failed on 
high limit, step 2 passed) 
Switches to the multi-step run page. 
Save currently edited file. 
:multi:speed 
<disc> 
:multi:speed? 
Sets or returns the measurement speed for the 
currently selected step. 
Send parameter: Max, Fast, Med, Slow 
Return parameter: 0(Max), 1(Fast), 2(Med), 3(Slow) 
Example: :multi:speed max (maximum speed) 
:multi:test <NR1> 
:multi:test? 
Selects or returns the step being edited. 
Parameter: 1 ~ 30 
Example: :multi:test 1(step 1 selected) 
Starts running multi-step measurements. 
Calibration command 
 
Performs open circuit trimming. 
Parameter: 1 (Spot trim), 2 (<10kHz), 3 (<100kHz), 
4 (All frequency) 
Example: :cal:oc-trim 4 (calibrate for all frequency) 
Performs short circuit trimming. 
Parameter: 1 (Spot trim), 2 (<10kHz), 3 (<100kHz), 
4 (All frequency), 5 (Rdc) 
Example: :cal:sc-trim 4 (calibrate for all frequency) 
Returns the result of the calibration performed. 
Parameter: 0 (fail), 1 (pass) 