PASS-FAIL MODE
67
Single-Step Test Run
Run in Absolute mode
1. Press F5 to select Absolute
measurement.
F 5
Abs %
2. Use the Left/Right key to move the
cursor to Hi/Lo value for editing.
HiLo 20.0mF0.00mF
3. Use the numerical keys
and unit keys to enter the
value. The editing value
appears at the bottom left
corner of the display.
Range
Speed Med
Auto
Lo Lim: 1.5_
Example
1.2m
m
Enter
Backspace
All clear
Clear
Increase
Decrease
Hi and Lo values are automatically swapped if
necessary
Hi and Lo Swapped
4. The display updates the Hi/Lo result
immediately. The result is pass if the bar stays
in the central area. The buzzer sounds
accordingly.