GRAPH MODE 
91 
Horizontal Scale Setting 
Set horizontal axis (Voltage) 
 
The X (horizontal) axis is selectable from Voltage 
and Frequency sweep. 
  When Voltage sweep is selected: measurement 
Frequency is fixed 
  When Frequency sweep is selected: 
measurement Voltage is fixed 
Select Drive level 
(Voltage) 
1.  Press the Up/Down key and move 
the cursor to Sweep. 
2.  If necessary, press the Left/Right 
key to change the sweep setting to 
Voltage (Drive Level). 
3.  Press the Up/Down key and move 
the cursor to Start. 
 
Use the numerical keys to enter the 
starting Voltage. 
10mV ~ 2V (AC≤3 MHz) 
10mV ~ 1V (AC>3 MHz)      *1mV step 
If a wrong unit is entered, the value is cancelled. 