Pass/Fail test (Single step)
Connect the fixture to the DUT.
Press the Menu key, then F5 (System).
Press the Up/Down arrow key to move
the cursor to Beep, then use the
Left/Right key to select setting (Off
recommended).
Press the Menu key, then F5 (System).
Press the Up/Down arrow key to move
the cursor to Average, then use the
numerical keys to enter an average
number (1-256). Press enter to confirm the
number.
Press the Menu key, followed by F1 (AC
measurement) or F2 (Rdc).
Press F4 (Show/Hide Scale) to show the
scale (or hide the circuit diagram)
6. Select measur-
ement item
Press F1 (First) and F2 (Second)
repeatedly to select the measurement
item.
7. Select series/
parallel circuit
If available, press F3 (Series/Parallel) to
select the equivalent circuit model.
8. Set
measurement
frequency
Press the Left/Right arrow key and move
the cursor to the Frequency. Use the
numerical and unit keys to set the
frequency.
9. Set
measurement
Voltage
Press the Left/Right arrow key and move
the cursor to the Voltage. Use the
numerical and unit keys to set the level.
10a. Select Single
measurement
Press the Sing/Rep key to select Single
(manual trigger) measurement. To trigger
measurement, press the Trig key.