EasyManua.ls Logo

GW Instek PSU 100-15 - Normal Function Settings

GW Instek PSU 100-15
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
101
MIN
Minimum internal resistance in ohms
MAX
Maximum internal resistance in ohms
Return parameter
<NR2>
Returns the internal resistance in ohms.
Example
SOUR:RES:LEV:IMM:AMPL 0.1
Sets the internal resistance to 100m.
[:SOURce]:VOLTage[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the voltage level in volts.
Syntax
[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
{<NR2>(V)|MINimum|MAXimum}
Query Syntax
[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Parameter
<NRf>
0~105% of the rated output voltage in
volts.
MIN
Minimum voltage level
MAX
Maximum voltage level
Return parameter
<NR2>
Returns the voltage level in volts
Example
SOUR:VOLT:LEV:IMM:AMPL 10
Sets the voltage level to 10 volts.
[:SOURce]:VOLTage[:LEVel]:TRIGgered
[:AMPLitude]
Set
Query
Description
Sets or queries the voltage level in volts when a
software trigger has been generated.
Syntax
[:SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]
{<NR2>(V)|MINimum|MAXimum}
Query Syntax
[:SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Parameter
<NR2>
0%~105% of the rated voltage output in
volts.

Table of Contents

Other manuals for GW Instek PSU 100-15

Related product manuals