Source Commands
[:SOURce]:CURRent[:LEVel][:IMMediate]
[:AMPLitude] ......................................................................... 95
[:SOURce]:CURRent[:LEVel]:TRIGgered
[:AMPLitude] ......................................................................... 96
[:SOURce]:CURRent:LIMit:AUTO .................................. 96
[:SOURce]:CURRent:PROTection:DELay ...................... 97
[:SOURce]:CURRent:PROTection[:LEVel] ..................... 97
[:SOURce]:CURRent:PROTection:STATe ...................... 98
[:SOURce]:CURRent:PROTection:TRIPped .................. 98
[:SOURce]:CURRent:SLEWrate:RISing ........................... 99
[:SOURce]:CURRent:SLEWrate:FALLing ....................... 99
[:SOURce]:MODE? ............................................................ 100
[:SOURce]:RESistance[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 100
[:SOURce]:VOLTage[:LEVel][:IMMediate]
[:AMPLitude] ....................................................................... 101
[:SOURce]:VOLTage[:LEVel]:TRIGgered
[:AMPLitude] ....................................................................... 101
[:SOURce]:VOLTage:LIMit:AUTO ................................ 102
[:SOURce]:VOLTage:LIMit:LOW .................................. 102
[:SOURce]:VOLTage:PROTection[:LEVel] .................. 103
[:SOURce]:VOLTage:PROTection:TRIPped ................ 103
[:SOURce]:VOLTage:SLEWrate:RISing ........................ 104
[:SOURce]:VOLTage:SLEWrate:FALLing .................... 104
[:SOURce]:CURRent[:LEVel][:IMMediate]
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps. For
externally set current levels (from the analog
control connector) the set current level is returned.
Syntax
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]
{<NR2>(A)|MINimum|MAXimum}
Query Syntax
[:SOURce]:CURRent[:LEVel][:IMMediate][:AMPLitude]
?
Parameter/Return
<NR2>
0~105% of the rated current output level.