PSU Series Programming Manual
112
:SYSTem:CONFigure:PROTection:RECovery
Set
Query
Description
Sets or queries how the OHP, FAN, AC-FAIL, and
SD alarms are cleared.
Syntax
Return Syntax
:SYSTem:CONFigure:PROTection:RECovery
{SAFE|AUTO}
:SYSTem:CONFigure:PROTection:RECovery?
Parameter
SAFE
The output is not turned on
automatically when the cause of the
alarm is fixed.
AUTO
The output is turned on automatically
when the cause of the alarm is fixed.
:SYSTem:CONFigure:MSLave
Set
Query
Description
Sets or queries the unit operation mode.
Note
This is the equivalent to the F-93 (Master/Slave
Configuration) power on configuration settings.
Syntax
Query Syntax
:SYSTem:CONFigure:MSLave { <NR1> }
:SYSTem:CONFigure:MSLave?
Parameter/
Return parameter
<NR1>
Description
0
Master/Local
1
Master/with 1 unit in Parallel (total:2
units)
2
Master/with 2 units in Parallel (total: 3
units)
3
Master/with 3 units in Parallel (total: 4
units)
4
Slave
Example
SYST:CONF:MSL 2
SYST:REB
Set to Master (with 2 slave units in Parallel) and
reboot the unit to active the setting.