Specifications
54400_spec.fm
Specifications
Specifications are subject to change without notice.
General
Display: LCD, 3½-digit measurement readout and six-character
alphanumeric scrolling text line.
Enclosure: IP62-rated with the gasketed door latched
Instrument Dimensions: 34.3 cm width x 41.9 cm height x 19.1 cm depth
(13.5 in. x 16.5 in. x 7.5 in.)
Mounting: Wall mount
Instrument Shipping Weight: 11.3 kg (25 lb)
Sample Requirements
Sample Flow Rate to Sample Conditioning: 200 to 500 mL/min
Inlet Pressure to Instrument:
1 to 5 psig; 1.5 psig is optimum. Exceeding 5 psig can cause sample
tubing failure.
Inlet Pressure to Sample Conditioning: 1.5 to 75 psig (with sample tube
level with the bottom of the instrument—see Figure 6 on page 19)
Sample Temperature Range: 5 to 40 °C (41 to 104 °F)
Inlet Fitting: at instrument—¼-inch OD polyethylene tubing with
quick-disconnect fitting
Drain Fitting: ½-inch hose barb
Sample Conditioning: Use provided sample conditioning
Reagent/Standard Requirements
Maximum Reagent Usage: one-half liter per month (each of the two
reagents)
Reagent Containers: High-density polyethylene (2) ½-liter bottles.
Reagent Containment: Reagent bottles are contained inside the analyzer
enclosure and are vented externally.
Electrical
Power Requirements: 100–115/230 V ac (selection switch inside
instrument); 95 VA, 50/60 Hz, 2.5 Amp fuse
Power Connection: Connection made by three wire barrier terminal block
through a ½-inch conduit hole in the case. Wire range: 12–18 AWG.
Installation Category: II

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach CL17 and is the answer not in the manual?

Hach CL17 Specifications

General IconGeneral
Power Requirements100-240 V AC, 50/60 Hz
Outlet PressureAtmospheric
Measurement MethodColorimetric DPD method
Sample Interval2.5 minutes
Operating Temperature5 - 40 °C (41 - 104 °F)
ParameterFree or Total Chlorine
Operating Humidity5 - 95% non-condensing
Sample Temperature0 to 40°C (32 to 104°F)
Sample Flow Rate200 to 500 mL/min
Power ConsumptionApproximately 50 W
Output4-20 mA, RS-232
RelaysTwo relays
EnclosureNEMA 4X
Resolution0.01 mg/L
Accuracy±5% or ±0.03 mg/L (whichever is greater)