Installation
19
Make wiring connections at the analyzer end as shown in Figure 19.
Figure 19 Analog Output Connections
3.4 Connecting/Wiring the Sensor Cable
The sensor cable is supplied with a keyed quick-connect fitting for easy
attachment to the controller, see Figure 20. Retain the connector cap to seal the
connector opening in case the sensor must be removed.
The 1720E sensor cable may be extended by a maximum of 7.62 m (25 ft),
see Replacement Parts and Accessories on page 55.
Modify the controller for sensor hard-wiring as follows:
1. Remove all power to the controller.
2. Open the controller cover.
Table 5 Output Connections at Terminal Block J6
Recorder Wires Circuit Board Position
Output 2 + 1
Output 2 – 2
Shield 3
Output 1 + 4
Output 1 – 5
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAY C
RELAY C
RELAY BRELAY A
RELAY A
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+ DATA
+ DATA
+ OUT 2
+ OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+ V
+ V
+ OUT 1
+ OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTS
ANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J5J5
NCNCNC
COMCOMCOM
NO
F1
F2
NONO
RELAY CRELAY CRELAY BRELAY ARELAY A
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+ DATA+ DATA
+ OUT 2+ OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+ V+ V
+ OUT 1+ OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANALOG OUTPUTSANALOG OUTPUTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J5J5
J6
1
2
3
4
5

Table of Contents

Related product manuals