42
System Startup
Note: If a dry verification is performed directly after a calibration, the measured value is
assigned as the expected value for future verifications (when using the dry verification
device with the same serial number). As long as the verification exists within the verification
history, the expected value will be retained. Otherwise, the expected value will be the
nominal labeled value associated with the dry verification device.
5.3.2 Calibration with StablCal
®
Before starting the calibration, read and apply the tips in section 5.3 on page 40.
6
OUTPUT MODE
Select ACTIVE, HOLD, or TRANSFER
a
Stop sample flow. Drain body and clean the body and bubble trap.
FILL CYL WITH 1 L DI WATER. REPLACE HEAD.
b
Measured reading (based on a gain of 1.0) displayed
c
(Remove head)
ADD 5 ML OF 4000 NTU FORMAZIN INTO CAL CYLINDER.
d
Measured reading (based on a gain of 1.0) displayed
e
GOOD CAL!
GAIN: X.XX
ENTER TO CONT
(to store)
f
Verify CAL?
(see Note below)
to verify
exit no verify
7
Select VERIFICATION type (begin at step 7 in section 5.4.1 on page 44 or
section 5.4.2 on page 45) or enter initials to complete calibration.
8
— RETURN SENSOR TO MEASURE MODE
9
Main Menu or Main Measurement Screen —
Step Select Menu Level/Instructions Confirm
Step Select Menu Level/Instructions Confirm
1
MAIN MENU —
2
SENSOR SETUP