Operation
33
4.5.2 Hold/Transfer Outputs
When cleaning or servicing the instrument, the analog outputs can be held at the
last measured values. To hold the output until released:
During calibration, the analog outputs can remain active, be held, or be
transferred to a preset mA value. When output hold or transfer is enabled during a
calibration, the hold or transfer is automatically released when calibration is
completed. See section 5.3, Sensor Calibration and Verification on page 40.
4.5.3 Release Outputs
Step Select Menu Level/Instructions Confirm
1
MAIN MENU —
2
TEST/MAINT
Enter Passcode if enabled
3
HOLD OUTPUTS
4
— SET OUTMODE
5
Select HOLD OUTPUTS or XFER OUTPUTS
6
SET CHANNELS
7
Select ALL or 1720E
8
ACTIVATION
9
— LAUNCH
10
Main Menu or Main Measurement Screen
reading will
flash
Step Select Menu Level Confirm
1
MAIN MENU —
2
TEST/MAINT
3
HOLD OUTPUTS
4
ACTIVATION