System Startup
45
5.4.2 Wet Verification
Before starting the verification, read and apply the appropriate tips in section 5.3
on page 40.
Step Select Menu Level/Instructions Confirm
1
MAIN MENU —
2
SENSOR SETUP
3
—CALIBRATE
4
VERIFICATION
5
— PERFORM VER
6
OUTPUT MODE
Choose ACTIVE, HOLD, TRANSFER
7
Select VER TYPE
Select WET
8
Enter Std Turbidity
a.
DRAIN AND CLEAN SENSOR BODY. ENTER TO CONT
b.
POUR STANDARD INTO CYL/BODY. PLACE HEAD ON. ENTER TO CONT
c.
Reading Displayed
(to accept)
d.
GOOD VER!
9
ENTER INITIALS
10
RETURN SENSOR TO MEASURE MODE
11
Main Menu or Main Measurement Screen —