38
Operation
RELAY SETUP (See section 4.6 on page 34 for expanded menu information.)
SELECT RELAY A, B, or C
SELECT SOURCE Select from none, any connected sensor, or the real time clock (RTC).
SET PARAMETER Press ENTER to select from the displayed parameters.
SET FUNCTION
Select from the available options to customize the relay functions. See section 4.6.1 on page 34 for
additional details.
SET TRANSFER Sets the relay to Energize or De-energize (user-selectable).
ACTIVATION Activate the relays from this menu (dependent on Function selected).
NETWORK SETUP (this menu appears only if a network card is installed in the controller)
MODBUS
ADDRESS
Highlight sc100 Analyzer, or either connected sensor then press ENTER to select. Choose a number
between 1 and 247 as the address (each source must have a different address) then press ENTER.
BAUD RATE Select a baud rate of 9600, 19200, 38.4K, 57.6K, or 115.2K. Default: 19200
STOP BITS Select 1 or 2 stop bits. Default: 1
MODBUS MODE Select RTU or ASCII. Default: RTU
DATA ORDER Select NORMAL or SWAPPED.
DISPLAY SETUP
ADJ CONTRAST
Use the UP and DOWN keys to increase or decrease the contrast, see section 4.3.2 on page 28.
Range = 0–50
LANGUAGE
The default is English. Choose from the available options to allow all menus to appear in the selected
language.
SET DATE/TIME
Use this menu to select the date format and to set the date and time (24-hour (military) format), see
section 4.3.4 on page 29.
SECURITY SETUP (Enter a 6-digit passcode)
SET PASSCODE
ENABLE Enables system security. See section 4.4.1 on page 31.
DISABLE Disables system security. See section 4.4.1 on page 31.
LOG SETUP (Not used for 1720E system. Enable datalogging from the sensor setup menu for 1720E)
DATALOG SETUP Set up datalogging of data and events. See section 4.7.1 on page 36.
ERROR HOLD MODE
HOLD OUTPUTS Holds outputs when unable to communicate with the sensor.
XFER OUTPUTS Goes to transfer state when unable to communicate with the sensor.
4.9.4 Test/Maint Menu
STATUS
Indicates the status of each relay and indicates which sensors are connected to the controller.
OUTPUT CAL
SELECT OUTPUT 1 or 2
Calibrate Analog Output by specifying values to correspond to 4 mA and 20 mA.
HOLD OUTPUTS
SET OUTMODE Choose Hold Outputs or Xfer Outputs.
SET CHANNELS Choose any individual attached sensor or all attached sensors to be held or transferred.
ACTIVATION Select Launch or Release.
4.9.3 System Setup Menu (continued)